Fresenius 2008 H, Dialysis Machine
Specification Dimensions (HxWxL) 50cm x 50cm x 114cm
Weight 155lbs.
Features
Modular component design.
Ultra precise volumetric UF control.
Blood flow rates of 0 to 600 ml/min.
Variable dialysate flow rates.
Proven reliability.
Automatic blood pressure monitoring (optional).
Sodium Variation up to 15%.
Computer interface (optional).
Specifications
Voltage 110, 220vac 50/60 hz
Current 12 amps max
Power 1350 watt @110 vac
Plug configuration NEMA 5-15 P
Environmental
Water pressure 1.0 to 7.0 bar
Water temperature 10.0 c to 25 c
Flow rate 900ml/min minimum
Drain height 46cm (18 in.) off floor max
Drain ID 4cm (1.5in.)
Air conditioning load 900 BTU/hr
Performance
Dialysate flow rate:
Minimum 300ml/min
Maximum 600ml/min
Standard 500ml/min
Ultra filtration control:
W / volumetric control 0-2000 ml/hr
W / TMP control 4500 ml/hr
Hot water rinse:
Temperature 85c
Display range 80c to 100c
Flow rate 300 ml/min
Time 20 min avg.
Venous pressure range -50 to +250 mm/hg
Arterial pressure range -150 to +250 mm/hg
TMP pressure range +60 to –540 mm/hg